MAbPac SCX-10 Columns
for the High-Resolution Separation and Analysis of Monoclonal Antibody Variants
The Thermo Scientific™ MAbPac™ SCX-10 column line has been expanded to include 3 and 5 μm particle
sizes for fast, high-resolution analysis as well as rapid separation
(RS) columns for bio-UHPLC separations. These smaller-particle-size
columns provide the same high resolution as the 10 μm columns, but with
significantly faster analysis time. In addition, salt or pH gradients
can be used to provide fast high-resolution, high-throughput analysis of
monoclonal antibody (MAb) variant samples.
The MAbPac SCX-10
columns are strong cation-exchange columns designed specifically for the
high-resolution, high-efficiency analysis of MAbs and associated
variants. The unique nonporous pellicular resin provides exceptionally
high resolving power, permitting the separation of MAb variants that
differ by as little as one charged residue. Hydrophobic interactions
with the resin are eliminated for very efficient peaks.
Features:
• Unequalled high-resolution separations
• Exceptionally high efficiency
• Highest resolution of MAb variants
• Unmatched column-to-column and lot-to-lot reproducibility
• Elimination of hydrophobic interactions
• Ideal for stability testing
The
MAbPac resin technology is the basis for the superior performance of
MAb variant analysis. The nonporous core particle provides high rates of
mass transfer, which results in high-efficiency separations. A
hydrophilic layer surrounds the polymeric beads, eliminating hydrophobic
interactions between proteins and the resin while contributing to
high-efficiency peaks. A proprietary grafted cation-exchange surface
provides pH selectivity control, resulting in high-resolution
separations. In addition, the MAbPac SCX-10 line of columns provides
consistent column-to-column and lot-to-lot reproducibility for MAb
variant separation results and is ideal for MAb therapeutic method
development and high-throughput screening.
The MAbPac SCX-10 RS columns are designed for use with the Thermo Scientific™ Dionex™ UltiMate™ 3000 BioRS system. These columns are packed in bioinert, PEEK-lined,
stainless steel column hardware suitable for operation up to 10,000
psi. These columns are higher efficiency, and therefore provide
higher-resolution separations.
MAbPac SCX-10 columns are designed
to address the regulatory requirements for biopharmaceutical
characterization. The consistent manufacturing processes ensure
reproducibility in methods development and data analysis.
The MAbPac SCX-10 column is complimentary to the industry-leading Thermo Scientific™ ProPac™ WCX-10 column for MAb variant analysis. The MAbPac SCX-10 column offers
alternative selectivity and provides higher resolution and efficiency
for variant analysis of most MAb samples than the ProPac WCX-10 column.
New MAbPac SCX Column Kits for Validation Now Available
-
Three MAbPac SCX-10, 4 × 250 mm columns from one lot of resin
-
Three MAbPac SCX-10, 4 × 250 mm columns, one each from three lots of resin
These
MAbPac SCX-10 column kits will facilitate the validation process of
MAbPac SCX-10 columns for MAb therapeutics development projects.
The MAbPac SCX-10, 10 μm, 9 × 250 mm column is now available for semipreparative applications of MAb variant separations.
MAbPac SCX-10 Monoclonal Antibody Analysis Column Specifications
|
Column Dimensions
|
10 μm: 4 × 250 mm, 4 × 150 mm, 4 × 50 mm, 2 × 250 mm
5 μm Analytical: 4 × 250 mm, 4 × 150 mm, 4 × 50 mm
5 µm Analytical RS: 4 × 250 mm, 4 × 150 mm
3 μm: 4 × 50 mm
|
Bead Size
|
Uniform 10 μm, 5 μm, 3 μm particles
|
Substrate
|
Highly crosslinked DVB media
|
Pellicular Layer
|
Proprietary hydrophilic
|
Functional Group
|
Sulfonic
|
Flow Rate Range:
|
0.2 to 2 mL/min for: 4 × 250 mm, 4 × 150 mm, 4 × 50 mm
0.1 to 0.5 ml/min for: 2 × 250 mm
|
Pressure Limit
|
3,000 psi for 10 μm columns: 4 × 250 mm, 4 × 150 mm, 4 × 50 mm
5,000 psi for Analytical 3 µm and 5 µm columns
7,000 psi for Analytical RS 5 µm columns
|
Temperature Range:
|
Ambient to 60 ºC
|
pH Range
|
2–12
|
Typical buffers:
|
Sodium or potassium salts of phosphate, MES
|
Solvents:
|
50% acetonitrile if needed for cleaning
|
Short Term Storage Solution (overnight):
|
Your low salt concentration eluent (pH 3–10)
|
Shipping Solution / Long Term Storage Solution:
|
20 mM MES (pH 5.6) + 60 mM NaCl– + 0.1% sodium azide
|
Detergent Compatibility
|
Nonionic, anionic, or zwitterionic detergents.
|
Analytical Columns
|
MAbPac SCX-10, 3µm, Analytical Column (4 x 50 mm)
|
077907
|
MAbPac SCX-10, 5µm, Analytical Column (4 x 50 mm)
|
078656
|
MAbPac SCX-10, 5µm, Analytical Column (4 x 150 mm)
|
085198
|
MAbPac SCX-10, 5µm, Analytical Column (4 x 250 mm)
|
078655
|
MAbPac SCX-10, 10 µm, Analytical Column (4 x 250mm)
|
074625
|
MAbPac SCX-10, 10µm, Analytical Column (4 x 150mm)
|
075602
|
MAbPac SCX-10HT, 10µm, Analytical Column (4 x 50mm)
|
075603
|
MAbPac SCX-10, 10µm, Analytical Column (2 x 250mm)
|
075604
|
MAbPac SCX-10, 10µm, 9 x 250 MM column
|
SP6866
|
MAbPac SCX-10, 10µm, Lot Select 3 Column Set (4 x 250 mm) (1 Lot of Resin)
|
SP6864
|
MAPac SCX-10, 10µm, Lot Select 3 Column Set (4 x 250 mm) (3 Lots of Resin)
|
SP6865
|
Guard Columns
|
MAbPac SCX-10G, 10µm, Guard Column (4 x 50 mm)
|
074631
|
MAbPac SCX-10G, 10µm, Guard Column (2 x 50 mm)
|
075749
|