The IonPac® AS20 hydroxide-selective anion-exchange column is specifically designed for determination of trace concentrations of perchlorate in drinking water, surface water, and groundwater matrices. The capacity and selectivity of the AS20 ensures that perchlorate can be quantified at low μg/L concentrations using suppressed conductivity detection, even in the presence of very high concentrations of chloride, carbonate, and sulfate. The new IonPac AS20 Capillary Column is packed with the same material as the equivalent standard bore version (providing the same performance as a 4 mm column) but requires only 1/100th the eluent flow rate. The capillary format provides the advantage of significantly reduced eluent consumption, which helps reduce operating costs.
The IonPac AS20 is the specified column in US EPA Method 314.1 (Confirmatory Method). The Cryptand C1 Concentrator Column is the specified concentrator column for sample preconcentration in that method.
Use the IonPac AS20 column with an RFIC™ system for optimum detection limits. Use the Anion Self-Regenerating Suppressor (ASRS® 300) or the Anion Capillary Electrolytic Suppressor (ACES 300) with the AS20 column for eluent suppression.
3.1 µeq (0.4 × 250 mm capillary column)77.5 µeq (2 × 250 mm analytical column)310 µeq (4 × 250 mm analytical column)0.06 µeq*(0.4 × 50 mm capillary guard column)1.5 µeq* (2 × 50 mm guard column)6 µeq* (4 × 50 mm guard column)*Guards are packed with a low-capacity microporous resin.